AI Triage Platform

DCB 0129, as manufacturer · GP Triage · 31 hazards, 26 suggested by Qyro

Controls 27Assumptions 4
27
Controls
21
Linked to the plan
3
Without an owner
5
Without evidence named
AllDesignTrainingProcessOther
ControlCategoryHazards protectedStatusOwnerEvidence requiredIn the Planner
Red-flag phrase libraryMaps everyday words for chest pain to the urgent pathwayDesignH-07 H-09In placeCSQuarterly review recordWatched monthly
Clinician review of urgent outcomesWithin 30 minutes, before the practice actsProcessH-07 H-03In placeNo ownerNot setAdd to the plan
Age check on the first screenUnder-16s are told to contact the practiceDesignH-03In placeZYRelease test recordWatched quarterly
Staff briefing on describing chest painTrainingH-07PlannedQLAttendance listNot yet
Certification against ISO 13485OtherNot linked to a hazardIn placeZYCertificateExpires 4 Mar 2027

Assumptions what the controls rely on.

Practices keep a duty clinician available during opening hoursRelied on by: Clinician review of urgent outcomes · confirmed by Riverside, 3 OctAccepted
Practices fall back to phone triage during an outageRelied on by: H-21 · not yet confirmed by any practiceAsk practices to confirm